 Purpose: Hydrophilicity is gaining increasing interest as a factor that might influence the osseointegration of dental implants. Therefore, in this study the dynamic wetting behavior of currently marketed dental titanium implants was analyzed by tensiometry, and its relationship to surface topography was examined.
Materials and Methods: Nine screw-type implant systems from eight manufacturers were evaluated. Dynamic water contact angles were analyzed by tensiometric multiloop Wilhelmy experiments (10 loops, 10 mm/min immersion speed). The wetted length (perimeter) of the immersed samples was estimated by three-dimensional picture profile measurement of the thread height of the respective implant screws. Wettability was quantified by first advancing contact angles. Additionally, static contact angles were determined using the sessile drop technique. All implant surfaces were characterized by scanning electron microscopy (SEM). Contact angle data were subjected to one-way analysis of variance followed by the Student t test.
Results: SEM revealed different types of surface morphology resulting from the different manufacturing processes. The first advancing mean contact angles of all implants ranged from 0 degrees (SLActive) to 138 degrees (OsseoSpeed), demonstrating statistically significant differences between implants. Because of kinetic hysteresis, initially hydrophobic implants became hydrophilic during following immersion loops.
Conclusions: The tensiometric method was used to compare wettability of dental implants. A range from fully wettable/superhydrophilic to virtually unwettable/hydrophobic was observed on the implant surfaces examined.


  Source : Journal of Oral & Maxillofacial Implant

  Full text

     
نوع مطالعه: پژوهشي | موضوع مقاله: پریودنتولوژی
دریافت: 1390/11/24 | انتشار: 1390/10/25
ارسال نظر درباره این مقاله
نام کاربری یا پست الکترونیک شما:

CAPTCHA


XML   English Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Rupp F, Scheideler L, Eichler M, Geis-Gerstorfer J. Wetting Behavior of Dental Implants. 3 2012; 6 (6)
URL: http://idai.ir/article-1-308-fa.html

Wetting Behavior of Dental Implants. 1. 1390; 6 (6)

URL: http://idai.ir/article-1-308-fa.html



بازنشر اطلاعات
Creative Commons License این مقاله تحت شرایط Creative Commons Attribution-NonCommercial 4.0 International License قابل بازنشر است.
دوره 6، شماره 6 - ( پریودنتولوژی 1390 ) برگشت به فهرست نسخه ها
جامعه اسلامی دندانپزشکان Islamic Dental Association of IRAN

کلیه حقوق سایت متعلق به جامعه اسلامی دندانپزشکان ایران است. تماس با ما

Persian site map - English site map - Created in 0.05 seconds with 37 queries by YEKTAWEB 4710