[Home ] [Archive]   [ فارسی ]  
Search in website

Advanced Search
..
:: Volume 9, Issue 9 (1-2012) ::
3 2012, 9(9): 0-0 Back to browse issues page
Cone beam CT image artefacts related to head motion simulated by a robot skull: visual characteristics and impact on image quality
R Spin-Neto * , J Mudrak , LH Matzen , J Christensen , E Gotfredsen , A Wenzel
Oral Radiology, Department of Dentistry, Aarhus University, Aarhus, Denmark
Abstract:   (1887 Views)

 Objectives: The aim of this study was to assess artefacts and their impact on cone beam CT (CBCT) image quality (IQ) after head motion simulated by a robot skull.

 Methods: A fully dentate human skull incorporated into a robot simulated pre-determined patient movements. Ten head motion patterns were selected based on the movement of the C-arm of the CBCT units (no motion as reference). Three CBCT units were used [a three-dimensional eXam (K) (KaVo Dental GmbH, Biberach, Germany), a Promax 3D MAX (P) (Planmeca Oy, Helsinki, Finland) and a Scanora® 3D (S) (Soredex Oy, Tuusula, Finland)]. Axial images were qualitatively assessed at three levels: mental foramen (MF), infraorbital foramen and supraorbital foramen, and artefacts characterized as stripe-like, double contours, unsharpness or ring-like. A 100 mm visual analogue scale (VAS) was used to quantitatively assess IQ. Cross-sectional images of the lower third molar and MF bilaterally were also evaluated by VAS. Four blinded examiners assessed the images.

 Results: For all units and motion patterns, stripe-like artefacts were the most common. The four observers agreed on the presence of at least one artefact type in 90% of the images. Axial images showed lower overall IQ after motion (VAS = 72.4 ± 24.0 mm) than reference images (VAS = 97.3 ± 2.6 mm). The most severe artefacts were seen at the MF level. For cross-sectional images, IQ was lowest after tremor. The mean IQ range was 74–89 and 57–90 for isolated (tilting, rotation and nodding) and combined (nodding + tilting and rotation + tilting) movements, respectively. IQ for MF was lower than for third molar for any movement except tremor.

 Conclusions: Head motion of any type resulted in artefacts in CBCT images. The impact on IQ depended on the region and level in the skull.


 Source: DentoMaxilloFacial Radiology

 Full Text

Keywords: cone beam computed tomography, artefacts, image quality
     
Type of Study: Research | Subject: Oral and Maxillofacial Radiology
Received: 2013/07/18 | Published: 2012/01/15
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA


XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Spin-Neto R, Mudrak J, Matzen L, Christensen J, Gotfredsen E, Wenzel A. Cone beam CT image artefacts related to head motion simulated by a robot skull: visual characteristics and impact on image quality. 3 2012; 9 (9)
URL: http://idai.ir/article-1-1374-en.html


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 9, Issue 9 (1-2012) Back to browse issues page
جامعه اسلامی دندانپزشکان Islamic Dental Association of IRAN

کلیه حقوق سایت متعلق به جامعه اسلامی دندانپزشکان ایران است. تماس با ما

Persian site map - English site map - Created in 0.06 seconds with 34 queries by YEKTAWEB 4714