The effect of temperature on the mechanical behavior of nickel-titanium orthodontic initial archwires
چکیده:   (1396 مشاهده)

 Objectives:

 To investigate and compare the characteristics of commonly used types of traditional and heat-activated initial archwires at different temperatures by plotting their load/deflection graphs and quantifying three parameters describing the discharge plateau phase.

  Materials and Methods:

 Forty-eight archwires of cross-sectional diameters ranging from 0.010 inches to 0.016 inches were obtained from seven different manufacturers. A modified three-point wire-bending test was performed on three analogous samples of each type of archwire at 55°C and 5°C, simulating an inserted archwire that is subjected to cold or hot drinks during a meal. For each resulting load/deflection curve the plateau section was isolated and the mean value of each parameter for each type of wire was obtained.

 Results:

 Permanent strain was exhibited by all wires tested at 55°C. Statistically significant differences were found between almost all wires for the three considered parameters when tested at 55°C and 5°C. Loads were greater at 55°C than at 5°C. Differences were also found between traditional and heat-activated archwires, the latter of which generated longer plateaus at 55°C, shorter plateaus at 5°C, and lighter mean forces at both temperatures. The increase in average force seen with increasing diameter tended to be rather stable at both temperatures.

 Conclusions:

 All nickel-titanium wires tested showed a significant change related to temperature in terms of behavior and force for both traditional and heat-activated wires. Stress under high temperatures can induce permanent strain, whereas the residual strain detected at low temperatures can be recovered from as temperature increases.


 Source: The Angel Orthodontists

 Full Text

واژه‌های کلیدی: NiTi archwire، Load deflection characteristics، Mechanical behavior
     
نوع مطالعه: پژوهشي | موضوع مقاله: ارتودنسی
دریافت: 1392/6/31 | انتشار: 1390/10/25
ارسال نظر درباره این مقاله
نام کاربری یا پست الکترونیک شما:

CAPTCHA


XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:


The effect of temperature on the mechanical behavior of nickel-titanium orthodontic initial archwires. 1. 1390; 1 (1 و 1)

URL: http://idai.ir/article-1-1971-fa.html



بازنشر اطلاعات
Creative Commons License این مقاله تحت شرایط Creative Commons Attribution-NonCommercial 4.0 International License قابل بازنشر است.
دوره 1، شماره 1 و 1 - ( مجموعه مقالات ارتودنسی 1390 ) برگشت به فهرست نسخه ها
جامعه اسلامی دندانپزشکان Islamic Dental Association of IRAN

کلیه حقوق سایت متعلق به جامعه اسلامی دندانپزشکان ایران است. تماس با ما

Persian site map - English site map - Created in 0.05 seconds with 37 queries by YEKTAWEB 4645