Dose indices in dental cone beam CT and correlation with dose–area product
چکیده:   (1811 مشاهده)

 Objectives: In the 2011 project “Safety and efficacy of a new and emerging dental X-ray modality (SEDENTEXCT)”, it was suggested that dose index (DI) and dose–area product (DAP) could be easily measured and used as diagnostic reference levels (DRLs), which would help in the management of radiation doses to patients in optimum exposure settings. Such indices could be directly related to effective dose. The purposes of this study, therefore, were to measure and calculate the DI and DAP in cone beam CT (CBCT) machines and to evaluate the correlation between the two.

 Methods: Dose measurements were performed on three-dimensional cone beam CT (3D-CBCT) machines [3D Accuitomo (J. Morita Mfg. Corp., Kyoto, Japan), Veraviewepocs (J. Morita Mfg. Corp.) and CS9300 (Carestream, New York, NY)] by exposing a cylindrical poly-methyl methacrylate (PMMA) phantom using a CT ionization chamber. These dose measurements were used for the calculation of Dose Indices 1 and 2, according to the methodology suggested by SEDENTEXCT. The DAP was measured using a DAP meter that was attached to the detector to cover the entire irradiated area.

  Results: The DI1 ranged from 53.6 mR to 216.6 mR, the DI2 ranged from 77.1 mR to 325.0 mR and the DAP ranged from 101.1 mGy cm2 to 457.9 mGy cm2, depending on the machines and exposure settings. Index 2 had a better correlation with the DAP than Index 1.

  Conclusions: The DIs and DAP proposed by SEDENTEXCT may be useful for establishing DRLs for dental CBCT machines however, further studies are necessary to determine which of these indices provide accurate dose estimates proportionally relating to the effective dose.


 Source: DentoMaxilloFacial Radiology

 Full Text

واژه‌های کلیدی: cone beam computed tomography، dose index، dose area product diagnostic reference levels
     
نوع مطالعه: پژوهشي | موضوع مقاله: رادیولوژی دهان، فک و صورت
دریافت: 1392/5/1 | انتشار: 1390/10/25
ارسال نظر درباره این مقاله
نام کاربری یا پست الکترونیک شما:

CAPTCHA


XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:


Dose indices in dental cone beam CT and correlation with dose–area product. 1. 1390; 9 (9)

URL: http://idai.ir/article-1-1418-fa.html



بازنشر اطلاعات
Creative Commons License این مقاله تحت شرایط Creative Commons Attribution-NonCommercial 4.0 International License قابل بازنشر است.
دوره 9، شماره 9 - ( رادیولوژی دهان، فک و صورت 1390 ) برگشت به فهرست نسخه ها
جامعه اسلامی دندانپزشکان Islamic Dental Association of IRAN

کلیه حقوق سایت متعلق به جامعه اسلامی دندانپزشکان ایران است. تماس با ما

Persian site map - English site map - Created in 0.05 seconds with 35 queries by YEKTAWEB 4710