Contrast curves of five different intraoral X-ray sensors: a technical note
|
|
|
|
چکیده: (1843 مشاهده) |
Objectives The aim of this article is to compare the contrast and resolution properties of five different sensors using a known technical target. Methods Stripe patterns with defined amounts of line pairs per mm (2.5-20 LP/mm) were recorded using five commercial digital sensors. Image data were analyzed using ImageJ and MatLab to calculate different contrast curves using logistic regression. Results The Dexis Platinum Sensor reached a calculated 10% contrast at 29.52 LP/mm. The Duerr VistaRay 6 Sensor reached a 10% contrast at 9.9 LP/mm. The 10% contrast was found at 18.8 LP/mm for the Duerr VistaRay 7. The Sirona Xios+ Sensor reached a calculated 10% contrast at 13.9 LP/mm. The Sirona Fullsize charge-coupled device (CCD) Sensor exhibited 10% contrast at 10.3 LP/mm. Conclusions The contrast transfer function assessment used in the study confirmed that the spatial frequency at 10% contrast was much lower than the theoretical resolution computed from the pixel size.
Source: Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology Full Text |
|
|
|
|
نوع مطالعه: پژوهشي |
موضوع مقاله:
رادیولوژی دهان، فک و صورت دریافت: 1392/5/27 | انتشار: 1390/10/25
|
|
|
|
|
ارسال نظر درباره این مقاله |
|
|