Object position and image magnification in dental panoramic radiography: a theoretical analysis
چکیده:   (1867 مشاهده)

 Objectives The purpose of our study was to investigate how image magnification and distortion in dental panoramic radiography are influenced by object size and position for a small round object such as a ball bearing used for calibration.

 Methods Two ball bearings (2.5 mm and 6 mm in diameter) were placed at approximately the same position between the teeth of a plastic skull and radiographed 21 times. The skull was replaced each time. Their images were measured by software using edge detection and ellipse-fitting algorithms. Using a standard definition of magnification, equations were derived to enable an object's magnification to be determined from its position and vice versa knowing the diameter and machine parameters.

 Results The average magnification of the 2.5 mm ball bearing was 1.292 (0.0445) horizontally and 1.257 (0.0067) vertically with a mean ratio of 1.028 (0.0322) standard deviations are in parentheses. The figures for the 6 mm ball bearing were 1.286 (0.0068), 1.255 (0.0018) and 1.025 (0.0061), respectively. Derived positions of each ball bearing from magnification were more consistent horizontally than vertically. There was less variation in either direction for the 6 mm ball bearing than the 2.5 mm one.

 Conclusions Automatic measurement of image size resulted in less variation in vertical magnification values than horizontal. There are only certain positions in the focal trough that achieve zero distortion. Object location can be determined from its diameter, measured magnification and machine parameters. The 6 mm diameter ball bearing is preferable to the 2.5 mm one for more reliable magnification measurement and position determination.


 Source: DentoMaxilloFacial Radiology

 Full Text

     
نوع مطالعه: پژوهشي | موضوع مقاله: رادیولوژی دهان، فک و صورت
دریافت: 1392/4/27 | انتشار: 1390/10/25
ارسال نظر درباره این مقاله
نام کاربری یا پست الکترونیک شما:

CAPTCHA


XML   English Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Devlin H, Yuan J. Object position and image magnification in dental panoramic radiography: a theoretical analysis. 3 2012; 9 (9)
URL: http://idai.ir/article-1-1365-fa.html

Object position and image magnification in dental panoramic radiography: a theoretical analysis. 1. 1390; 9 (9)

URL: http://idai.ir/article-1-1365-fa.html



بازنشر اطلاعات
Creative Commons License این مقاله تحت شرایط Creative Commons Attribution-NonCommercial 4.0 International License قابل بازنشر است.
دوره 9، شماره 9 - ( رادیولوژی دهان، فک و صورت 1390 ) برگشت به فهرست نسخه ها
جامعه اسلامی دندانپزشکان Islamic Dental Association of IRAN

کلیه حقوق سایت متعلق به جامعه اسلامی دندانپزشکان ایران است. تماس با ما

Persian site map - English site map - Created in 0.06 seconds with 35 queries by YEKTAWEB 4710